Intelect Transport 2 Combo
Intelect Transport 2 Combo is a two-channel electrotherapy, ultrasound therapy combo system providing Chattanooga product reliability and value. It’s lightweight and durable design with battery operation option provides the versatility to treat patients in the clinic or on the field. Its intuitive functionality provides ease of use and confidence in effective therapy.
- Electrical Stim: 5 waveforms (Interferential, Premod, Russian, High Volt, VMS)
- Dual Frequency Ultrasound: 1 MHz and 3.3 MHz frequencies with pulsed or continuous duty cycles (10%, 20%, 50%, and 100%). Comes standard with a 5cm2 ultrasound applicator
Other Features of the Intelect Transport 2
- Large Color LCD Screen: An easy to see high resolution screen provides real time information on the progress of the individual treatment.
- Portability: Its lightweight and durable design coupled with an ergonomically designed handle make transporting the devise easy to treat patients whenever and wherever.
- Optional Cart: Complete with large wheels and casters, magnetic fixation of the device to the cart for convenience and easy and secure attachment and detachment, and sufficient storage drawers make this optional cart a convenient and functional treatment, storage, and transport system. Dimensions: W-48cm, D-52cm, H-96cm, (108.5cm with device on cart). Weight: 22.27 lbs (10.1 kg)
- Ultrasound: Improved coupling detection and monitoring for more effective and efficient treatments.
- VMS Waveform:Chattanooga's proprietary VMS waveform achieves stronger muscle contractions and less neural fatigue while still providing a more comfortable experience for the patient. (
- Improved Battery: Allows for reduced need for excess cables and provides all day use in portable settings.
- Warranty: During the two-year warranty period from the date of delivery of the product to the end customer, defects will be remedied at no charge to the customer upon the customer furnishing adequate proof that the defect is due to defects in material or workmanship. The warranty period for accessories is 90 days. Accessories consist of Lead Wires and Electrodes. The warranty period for the Therapy System Cart and Ultrasound Applicators is one year (12 months).
Click HERE to view Intelect Transport 2 Combo Brochure
Click HERE to view Intelect Transport 2 Combo User’s Manual
Intelect Transport 2 Standard Accessories |
|
SKU/Part Number |
Description |
4738 |
Intelect Transport 2 Combo |
14679 |
Mains Power Cord |
47335 |
Transport 2 Ultrasound Applicator 5cm2 |
48384 |
Transport 2 User Manual |
40-0097 |
Leadwire Assy, Stim, CH1, STD |
40-0098 |
Leadwire Assy, Stim, CH1, STD |
42192 |
Dura-Stick Plus 5 cm (2”) Foam, Round Electrodes |
Intelect Transport 2 Optional Accessories |
|
SKU/Part Number |
Description |
70010 |
Stim CH1/2 Leadwire Kit STD (For Combo Device Only) |
10648 |
Nylatex Wrap (2 per pack) |
79967 |
6 x 8 cm (2.5 x 3 in) Carbon Electrodes (4 pk) |
79970 |
Sponge Pocket 9 x 10.5cm (4 pk) |
72852 |
3 in (8cm) Black Rubber Carbon Electrodes (2 pk) |
72853 |
3 in (8cm) Red Rubber Electrodes (2 pk) |
47334 |
Transport 2 Ultrasound Applicator 2cm2 |
47336 |
Transport 2 Ultrasound Applicator 10cm2 |
14-1086 |
Battery |
15-1136 |
Cart |
Indications for Use – Electrotherapy |
- Relaxation of muscle spasms
- Prevention or retardation of disuse atrophy
- Increasing local blood circulation
- Muscle re-education
- Maintaining or increasing range of motion
- Symptomatic relief of chronic, intractable pain
- Post-traumatic acute pain
- Post-surgical acute pain
Indications for Use – Ultrasound |
Relief of pain, muscle spasms, and joint contractures associated with a wide variety of musculoskeletal injuries and conditions.
- “itis” conditions
- Muscle spasms
- Musculoskeletal syndromes/conditions
- Muscle & ligament strains and sprains
- Trigger points
- Relief of sub-chronic and chronic pain and joint contractures resulting from capsular tightness and scaring
Technical Specifications:
Mains Power |
100-240 VAC, 50/60 Hz, 1.0 A |
Weight |
4.9 lbs (2.2 kg) |
Shipping Weight |
8.5 lbs (3.9 kg) |
Dimensions |
13.6"L x 9.8"W x 4.7"H (34.6 cm x 25 cm x 12 cm) |
Electrical Safety Class |
Class 1, Electrotherapy Type BF, Ultrasound Type B |
Safety Tests |
IEC 60601-1, ISO 10993-1, IEC 60601-2-5, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-6, IEC 62366-1, IEC 62304, IEC 62133, ISO 14971 |
SYSTEM SPECIFICATIONS
Power
- Input: 100-240 VAC, 1.0 to 0.42 A, 50/60 Hz 100 W Max
- Electrical Class: Class 1
- Mode of Operation: Continuous
Output Power
ULTRASOUND SPECIFICATIONS |
|
Frequency |
1 MHz; 3.3 MHz |
Duty Cycles |
10%, 20%, 50%, Continuous |
Pulse Repetition Rate |
100 Hz |
Pulse duration |
1 -5 ms Max (ON): 5 ms Min (OFF): 5ms |
Peak to Average Ratio |
1:1, at 50% Duty Cycle 4:1, at 20% Duty Cycle 9:1, at 10% Duty Cycle |
Beam Nonuniformity Ratio |
<5.1 |
Beam Type |
Collimating |
Treatment Time |
1 to 30 minutes |
WAVEFORMS
IFC (Interferential) Traditional (4 Pole) |
Interferential Current is a medium frequency waveform. Current is distributed through two channels (four electrodes). The currents cross each other in the body at the area requiring treatment. The two currents interfere with each other at this crossing point, resulting modulation of the intensity (the current intensity increases and decreases at a regular frequency).
Output Mode |
Electrodes |
Available on Channels |
1, 2 |
Treatment Time |
1-60 Minutes |
Mode Selection |
CC or CV |
Output Intensity |
0-100 mA (CC) |
Beat Frequency |
1-200 Hz |
Carrier Frequency |
2000-10,000 Hz |
Sweep Time |
15 sec |
Sweep Low Beat Frequency |
1-199 Hz |
Sweep High Beat Frequency |
2-200 Hz |
Scan Percentage |
Static, 40%, 100%, Manual |
IRMS |
0-78mA |
DC component |
No |
VMS |
VMS is a symmetrical biphasic waveform with a 100 µsec interphase interval. Because the pulse is relatively short, the waveform has a low skin load, making it suitable for applications requiring high intensities, such as in muscle re-education protocols.
Output Mode |
Electrodes |
Available on Channels |
1, 2 |
Treatment Time (Stim) |
1-60 minutes |
Treatment Time (Combo) |
1-30 minutes |
Mode Selection |
CC or CV |
Output Intensity |
0-140 mA (CC) 0-140 V (CV) |
|
|
Channel Mode |
Single, Reciprocal, Co-Contract |
Cycle Time |
Continuous 5/5, 4/12, 10/10, 10/20, 10/30, 10/50 |
Frequency |
1 – 200 pps |
Phase Duration |
20-400 μsec |
Ramp |
0.5 – 5 seconds |
Set Intensity |
Individual/both Channel Intensity, Setting in Reciprocal and Co-Contract Modes |
IRMS |
0-50mA |
DC Component |
No |
IFC Premodulated (Traditional 2-Pole) |
Premodulated Current is a medium frequency waveform. Current comes out of one channel (two electrodes). The current intensity is modulated: it increases and decreases
at a regular frequency (the Amplitude Modulation Frequency).
Output Mode |
Electrodes |
Available on Channels |
1, 2 |
Treatment Time (Stim) |
1-60 minutes |
Treatment Time (Combo) |
1-30 minutes |
Mode Selection |
CC or CV |
Output Intensity |
0-100 mA (CC) 0-100 V (CV) |
Beat Fixed (Sweep Off) |
1 – 200 Hz |
Cycle Time |
Continuous, 5/5, 4/12, 10/10, 10/20, 10/30, 10/50 |
Carrier Frequency |
2,500 Hz |
Sweep Low Beat Frequency |
1 – 199Hz |
Sweep High Beat Frequency |
2 – 200Hz |
IRMS |
0-55mA |
DC Component |
No |
Russian |
Russian Current is a sinusoidal waveform, delivered in bursts or series of pulses. This method was claimed by its author (KOTA) to produce maximal muscle re-education effects without significant discomfort to patient.
Output Mode |
Electrodes |
Available on Channels |
1, 2 |
Treatment Time |
1-60 minutes |
Mode Selection |
CC or CV |
Output Intensity |
0-100 mA (CC) 0-100 V (CV) |
Channel Mode |
Single, Reciprocal, Co-Contract |
Burst Frequency |
20 – 100 bps |
Carrier Frequency |
2,500Hz |
Cycle Time |
Continuous, 5/5, 4/12, 10/10, 10/20, 10/30, 10/50 |
Duty Cycle |
10%, 20%, 30%, 40%, 50% |
Ramp |
0.5 – 5 seconds |
IRMS |
0-39mA |
DC Component |
No |
High Voltage Pulsed Current (HVPC) |
The High Voltage Pulsed Current (HVPC) has a very brief pulse duration characterized by two distinct peaks delivered at high voltage. The waveform is monophasic (current flows in one direction only). The high voltage causes a decreased skin resistancemaking the current comfortable and easy to tolerate.
Output Mode |
Electrodes |
Available on Channels |
1, 2 |
Treatment Time (Stim) |
1-60 minutes |
Treatment Time (Combo) |
1-30 minutes |
Mode Selection |
CV |
Output Intensity |
0 – 500 V (CV) |
Cycle Time |
Continuous, 5/5, 4/12, 10/10, 10/20, 10/30, 10/50 |
Display |
Volts or Peak Current |
Frequency |
10 – 120 pps |
Polarity |
Positive or Negative |
Ramp |
0.5 – 5 seconds |
Sweep Frequency |
Continuous, 80/120, 1/120, 1/10 |
IRMS |
0-45mA |
DC Component |
0 – 1.5mA |